8-64
Operation Manual
ξξ Dist 20
Traces the cardiac circumference and then draw the cardiac axis. The system automatically draws 20
straight lines perpendicular to the axis and calculates its volume.
NOTE:
ξX
MR dp/dt can be measured only at 1-3m/s.
ξX
In Dual 2D Mode, two images can be viewed simultaneously.
ξX
For RVAWd, RVIDd, RVAWs, and RVIDs, see the LV measurement method.
ξX
MPA Diam, RPA Diam, and LPA Diam are measured under Aortic Valve Level in Parasternal
Short Axis.
ξX
C Mode is mainly used for measuring reverse cardiac blood flow.
ξX
As PISA-Radius or PISA-Alias Vel. measurements require Velocity values, you have to select
color display for Velocity or Vel + Var in C Mode. For more information, refer to the βColor Doppler
Modeβ section in Chapter 7. βDiagnosis Modeβ.
ξX
Tissue Doppler can be measured in TDI Mode.