Chapter 4 Diagnosis Modes
3. Tap Start/Stop to end acquiring the panoramic image. The touch screen will be switched to the
Panoramic Review screen.
Tips!
Cautions for Acquiring a Panoramic Image
X
When scanning a curved surface, ensure that the scan surface and the contact surface of the
probe are always at right angles.
X
Moving in the opposite direction while acquiring an image erases the previous saved frames
and saves new frames.
X
The image quality may deteriorate if the contact surface of the probe loses contact with the
scan surface.
X
If the scan speed is fast or the contact surface of the probe changes in angle, artifacts may occur.
[Figure 4.4 Panoramic Ready -Touch Screen]
Reviewing a Panoramic Image
NOTE:
X
You can perform basic measurements using the Caliper button. But other functions on the
control panel are not available.
X
L/R Flip, U/D Flip, and Magnifying are available only when Layout is set to Full Screen.