5-15
Chapter 5 Measurements and Calculations
D Trace
This is a basic measurement that is available in Spectral Doppler Mode only. You can specify a point in
a Spectral Doppler image and trace a curve from that point to calculate the velocity, integral value and
average velocity of blood ow.
1.
Select t
he Soft Menu dial-bu
tton
[1] D Trace. “D Trac
e” is displayed in the user information area.
2.
Trace a curve. The method for measuring a curve is the same as in “2D Line Trace.
3.
When the measurement is nished, its result is shown on the screen.
X
PSV: Peak Systolic Velocity
X
EDV: End Diastolic Velocity
X
PI: Pulsatility Index
X
RI: Resistivity Index
X
S/D: Ratio of PSV to EDV
X
Vmax: Max Velocity
X
Vmean: Mean Velocity
X
PGmax: Max Pressure Gradient
X
PGmean: Mean Pressure Gradient
X
VTI: Velocity Time Integral
X
PHT: Pressure Half Time
X
Acc: Acceleration
X
AccT: Acceleration Time
X
Dec: Deceleration
X
DecT: Deceleration Time
The equations used for D Trace measurement are as follows:
X
X
X
X
NOTE: If Application is set to Cardiac at Utility > Measure Setup > General > Caliper, the results
including Vmax, PGmax, PGmean, VTI, PHT, Acc, AccT, Dec and DecT will be shown on the screen.