Samsung SONOACE R7 Manuale Utente Pagina 199

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 224
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 198
9-39
Chapter 9 Probes
Disinfection
CAUTION:
Plastic biopsy kits can only be disinfected using a chemically compatible cold-disinfectant.
Disinfection by autoclaving or by using gas or radiation will cause damage to these parts.
Please refer to the user manual of the disinfectant for storage, use, and disposition of the disinfectant.
1. Check the disinfection duration (generally 10 hours) and temperature of the disinfectant.
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method used.
3. Inspect the components for damage such as cracks, rust or breakage. If damage is evident, discontinue
use of the biopsy kit and contact Samsung Medison
Customer Service Department.
Vedere la pagina 198
1 2 ... 194 195 196 197 198 199 200 201 202 203 204 ... 223 224

Commenti su questo manuale

Nessun commento