5-24
Vesl. Dist. (Vessel Distance)
Measure the width of a blood vessel and calculate TAMV and Volume Flow.
Stenosis Measurement
You can measure the stenosis of each blood vessel system by measuring and calculating an area or dis-
tance.
% StA (Stenosis Area)
Measure the area of the inner and outer walls of a blood vessel.
1. Select the %StA menu and the rst cursor will appear in 2D Mode.
2. Measure the area of the vessel’s outer wall using the Circ / Area measurement method.
3.
When the second cursor appears, measure the area of the vessel’s inner wall under stenosis.
%Stenosis Area. = (Outer Area – Inner Area) / Outer Area × 100
% StD (Stenosis Distance)
Measure the diameter of a blood vessel.
1. Select the %StD menu and the rst cursor will appear in 2D Mode.
2. Measure the total diameter of a blood vessel using the Distance measurement method.
3.
When the second cursor appears, measure the diameter of the vessel’s inner wall under stenosis.
%Stenosis Dist. = (Outer Distance – Inner Distance) / Outer Distance × 100
Heart Rate Measurement
HR (Heart Rate)
You can calculate heart rates for a certain period of time.
1. Select HR in the Measurement menu. A bar appears allowing you to specify a measurement area.
2. Specify the measurement range.
Place the bar at a desired position with the trackball, and press the Set button.
3. The system measures the heart rate within the specied range automatically. The measurement
results are displayed on the screen.