Samsung VP-D310 Specifiche Pagina 331

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 503
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 330
OWNER’S GUIDE  CONDITIONERS, DAQ | 13331
In the quarter bridge 4-wire mode the DAQP-STG internally adjusts its excitation so that on the gage the
resistor terminates exactly on the half of the excitation voltage. All wire resistances are compensated.
DAQP-STG Resistance, RTD 2-wire and 4-wire
For resistance and RTD mode, 4-wire connection is recommended (2-wire connection will not compensate
wire resistance).
DAQP-STG - Other measurement modes and hook-ups
Voltage and microvolt measurement signal connection
CAUTION: If the excitation is not used for sensor supply it has to be deactivated by setting it to 0 V. This will
internally connect the IN- to the GNDisolated to improve the common mode rejection.
Vedere la pagina 330
1 2 ... 326 327 328 329 330 331 332 333 334 335 336 ... 502 503

Commenti su questo manuale

Nessun commento