Acceptance criteria:
¾ A correlation coefficient (r) of the calibration curve must be > 0.9900.
¾ The back-calculated concentration of the lowest calibration standard (CS1) must be
within 80-120% of its theoretical concentration.
¾ The back-calculated concentrations of all other calibration standards must be within 85-
115% of their theoretical concentrations.
¾ The curve must contain at least 75% of the calibration standards for evaluation of curve
fitting.
¾ No two adjacent (or consecutive) calibration standards can be rejected.
¾ Both replicates each of LLOQ and ULOQ standard cannot be rejected, either of the two
replicate must meet the acceptance criteria.
E. Accuracy:
1. Within-batch or intra-batch accuracy:
¾ Calibration standards and six replicates each of LLOQ, LQC, MQC and HQC samples
were processed and analyzed as per the procedure described in sample preparation
section.
¾ % nominal concentration of back-calculated value for LLOQ, LQC, MQC and HQC,
analyzed in single analytical batch was calculated, as per formula to determine within-
batch or intra-batch accuracy.
2. Between -batch or inter-batch accuracy:
¾ Calibration standards and six replicates each of LLOQ, LQC, MQC and HQC samples
were processed and analyzed as per the procedure described in sample preparation
section.
¾ Five different batches on different days were performed.
¾ % nominal concentration of back-calculated value for LLOQ, LQC, MQC and HQC
samples, analyzed on five different batches on different days was calculated.
Acceptance criteria:
¾ The back calculated concentrations of all QC samples (LQC, MQC, and HQC) must be
within 85-115% of their nominal concentration except at LLOQ sample where it should
not deviate by more than 80-120% of its nominal concentration.
¾ At least 67% quality control samples must fall within above-mentioned criteria at each
LLOQ, LQC, MQC, and HQC levels.